Abstract:

Background In-hospital cardiac arrest has a poor prognosis despite active electrocardiography monitoring. The initial rhythm of approximately 25% of in-hospital cardiopulmonary resuscitation (CPR) events is pulseless ventricular tachycardia/ventricular fibrillation (VT/VF). Early defibrillation is an independent predictor of survival in CPR events caused by VT/VF. The automated external cardioverter defibrillator (AECD) is a device attached by pads to the chest wall that monitors, detects, and within seconds, automatically delivers electric countershock to an appropriate tachyarrhythmia. Study Objectives • To evaluate safety of AECD monitoring in hospitalized patients. • To evaluate whether AECDs provide earlier defibrillation than hospital code teams. Methods The study is a prospective trial randomizing patients admitted to the telemetry ward to standard CPR (code team) or standard CPR plus AECD monitoring (PowerHeart CRM). The AECD is programmed to deliver one 150 J biphasic shock to patients in sustained VT/VF. Data is collected using the Utstein criteria for cardiac arrest. The primary endpoint is time-to-defibrillation; secondary outcomes include neurological status and survival to discharge, with 3-year follow-up. Results To date, 192 patients have been recruited in the time period between 10/10/2006 to 7/20/2007. A total of 3,655 hours of telemetry data have been analyzed in the AECD arm. The AECD has monitored ambulatory telemetry patients in sinus rhythm, sinus tachycardia, supraventricular tachycardia, atrial flutter or fibrillation, with premature ventricular complexes and non-sustained VT without delivery of inappropriate shocks. One patient experienced sustained VT during AECD monitoring, who was successfully defibrillated (17 seconds after meeting programmed criteria). There are no events to report in the control arm. The patient survived the event without neurological complications. During the same time period, mean time to shock for VT/VF cardiac arrest occurring outside the telemetry ward was 230 ± 50 seconds.Conclusion AECD monitoring is safe and likely results in earlier defibrillation than standard telemetry monitoring. Trial Registration National Institutes of Health registration ID: NCT00382928

Copyright information:

© 2008 Ali et al; licensee BioMed Central Ltd.

This is an Open Access work distributed under the terms of the Creative Commons Attribution 2.0 Generic License (http://creativecommons.org/licenses/by/2.0/).

Creative Commons License

Export to EndNote