Background: Intentional laceration of the anterior mitral leaflet (LAMPOON) is an effective adjunct to transcatheter mitral valve replacement that prevents left ventricular outflow tract (LVOT) obstruction. To date, LAMPOON has been performed in over 150 patients using a retrograde approach that can be technically challenging. A modified antegrade transseptal technique may simplify the procedure. Methods: Antegrade LAMPOON was developed and tested in nonsurvival pig experiments. Thereafter, antegrade LAMPOON was performed in patients at prohibitive risk of LVOT obstruction. Clinical, procedural, and angiographic details were abstracted from medical records of their index procedure, and were compared with findings in comparable patients at risk of fixed-LVOT obstruction in the LAMPOON investigational device exemption trial. Results: Eight patients at risk of fixed LVOT obstruction underwent antegrade LAMPOON. Leaflet traversal and laceration were technically successful in all. There were no cases of clinically significant LVOT obstruction (mean LVOT gradient at discharge: 5.4±1.4 mm Hg). One patient suffered a ventricular wire perforation, unrelated to the antegrade LAMPOON technique, and did not survive to discharge. At the time of discharge, no patients had an increase of >10 mm Hg in LVOT gradient compared with baseline. Procedure times (from traversal to transcatheter mitral valve replacement) were shorter, compared with the retrograde technique in the LAMPOON investigational device exemption trial (39±09 versus 65±35 minutes). All patients survived (8/8, 100%) the procedure, and 7/8 (88%) survived to 30 days, similar to subjects in the LAMPOON investigational device exemption trial. Conclusions: Antegrade LAMPOON is an effective, reproducible, and simplified strategy to lacerate the anterior leaflet before transcatheter mitral valve replacement. The authors recommend the technique as the new standard for LAMPOON.
OBJECTIVES
The aim of this study was to test the hypothesis that transcatheter electrosurgery might allow intentional detachment of previously placed MitraClip(s) from the anterior leaflet to recreate a single mitral orifice for transcatheter mitral valve implantation (TMVI), leaving the retained MitraClip(s) securely fastened to the posterior leaflet and without interfering with the mitral bioprosthesis.
BACKGROUND
Patients with severe mitral regurgitation or stenosis despite edge-to-edge mitral repair with the MitraClip typically have few therapeutic options because the resultant double orifice precludes TMVI. Transcatheter electrosurgery may allow detachment of failed MitraClip(s) from the anterior leaflet to recreate a single orifice for TMVI.
METHODS
This was a single-center, 5-patient, consecutive, retrospective observational cohort. Patients underwent transcatheter electrosurgical laceration and stabilization of failed MitraClip(s) to recreate a single orifice, leaving the MitraClip(s) securely fastened to the posterior leaflet. Subsequently, patients underwent TMVI with an investigational device, the Tendyne mitral bioprosthesis, on a compassionate basis. Patients were followed up to 30 days.
RESULTS
MitraClip detachment from the anterior leaflet and Tendyne implantation were successful in all patients. All patients survived to discharge. All patients were discharged with grade 0 central mitral regurgitation. Two patients had moderate perivalvular mitral regurgitation that did not require reintervention. During the follow-up period of 30 days, there were no deaths, cases of valve dysfunction, or reintervention. There was no evidence of erosion or bioprosthetic valve dysfunction attributable to the retained MitraClip(s) still attached to the posterior leaflet.
CONCLUSIONS
Transcatheter electrosurgical detachment of failed MitraClips from the anterior leaflet followed by TMVI is technically feasible and safe at 30 days. Longer term study is needed to determine the clinical benefit of this approach and new algorithms for TMVI sizing following electrosurgical laceration and stabilization of a failed MitraClip to avoid perivalvular leak.
The SAPIEN 3 is the only transcatheter heart valve commercially available for compassionate transcatheter mitral valve replacement in patients with previous mitral surgical rings and mitral annular calcification (valve in ring [VIR] and valve in mitral annular calcification [VIM]). Reported outcomes have been inconsistent or poor. The review provides an overview of the authors’ approach to achieve largely consistent results despite the intrinsic limitations of SAPIEN 3 VIM and VIR. The approach includes bedside modifications of the valve implant, the delivery system, and of the cardiac substrate itself. Until purpose-built devices are readily available, VIR and VIM procedures will require aggressive multidisciplinary cooperation, meticulous planning and execution, and postprocedure management by experienced, high-volume operators.
Objectives: The purpose of this study was to evaluate tip-to-base intentional laceration of the anterior mitral leaflet to prevent left ventricular outflow tract obstruction (LAMPOON) in patients undergoing transcatheter mitral valve replacement (TMVR) in annuloplasty rings or surgical mitral valves. Background: LAMPOON is an effective adjunct to TMVR that prevents left ventricular outflow tract obstruction (LVOTO). Laceration is typically performed from the base to the tip of the anterior mitral leaflet. A modified laceration technique from leaflet tip to base may be effective in patients with a prosthesis that protects the aortomitral curtain. Methods: This is a multicenter, 21-patient, consecutive retrospective observational cohort. Patients underwent tip-to-base LAMPOON to prevent LVOTO and leaflet overhang, or therapeutically to lacerate a long anterior mitral leaflet risking or causing LVOTO. Outcomes were compared with findings from patients in the LAMPOON investigational device exemption trial with a prior mitral annuloplasty. Results: Twenty-one patients with a annuloplasty or valve prosthesis–protected mitral annulus underwent tip-to-base LAMPOON (19 preventive, 2 rescue). Leaflet laceration was successful in all and successfully prevented or treated LVOTO in all patients. No patients had significant LVOTO upon discharge. There were 2 cases of unintentional aortic valve injury (1 patient underwent emergency transcatheter aortic valve replacement and 1 patient underwent urgent surgical aortic valve replacement). In both cases, the patients had a supra-annular ring annuloplasty, and the retrograde aortic guiding catheter failed to insulate the guidewire lacerating surface from the aortic root. All patients survived to 30 days. Compared with classic retrograde LAMPOON, there was a trend toward shorter procedure time. Conclusions: Tip-to-base laceration is a simple, effective, and safe LAMPOON variant applicable to patients with an appropriately positioned mitral annular ring or bioprosthetic valve. Operators should take care to insulate the lacerating surface from adjacent structures.