Characterization of 250 MeV Protons from the Varian ProBeam PBS System for FLASH Radiation Therapy.

Downloadable Content

Persistent URL
Last modified
  • 06/25/2025
Type of Material
Authors
    Serdar Charyyev, Stanford UniversityChih-Wei Chang, Department of Radiation Oncology and Winship Cancer Institute, Emory University, Atlanta, GA, USA.Mingyao Zhu, Emory UniversityLiyong Lin, Emory UniversityKatja Langen, Emory UniversityAnees Dhabaan, Emory University
Language
  • English
Date
  • 2023
Publisher
  • Elsevier Inc
Publication Version
Copyright Statement
  • ©2023 The Author(s)
License
Final Published Version (URL)
Title of Journal or Parent Work
Volume
  • 9
Issue
  • 4
Start Page
  • 279
End Page
  • 289
Grant/Funding Information
  • The authors have no funding to disclose.
Abstract
  • Shoot-through proton FLASH radiation therapy has been proposed where the highest energy is extracted from a cyclotron to maximize the dose rate (DR). Although our proton pencil beam scanning system can deliver 250 MeV (the highest energy), this energy is not used clinically, and as such, 250 MeV has yet to be characterized during clinical commissioning. We aim to characterize the 250-MeV proton beam from the Varian ProBeam system for FLASH and assess the usability of the clinical monitoring ionization chamber (MIC) for FLASH use. We measured the following data for beam commissioning: integral depth dose curve, spot sigma, and absolute dose. To evaluate the MIC, we measured output as a function of beam current. To characterize a 250 MeV FLASH beam, we measured (1) the central axis DR as a function of current and spot spacing and arrangement, (2) for a fixed spot spacing, the maximum field size that achieves FLASH DR (ie, > 40 Gy/s), and (3) DR reproducibility. All FLASH DR measurements were performed using an ion chamber for the absolute dose, and irradiation times were obtained from log files. We verified dose measurements using EBT-XD films and irradiation times using a fast, pixelated spectral detector. R90 and R80 from integral depth dose were 37.58 and 37.69 cm, and spot sigma at the isocenter were σx = 3.336 and σy = 3.332 mm, respectively. The absolute dose output was measured as 0.343 Gy*mm2/MU for the commissioning conditions. Output was stable for beam currents up to 15 nA and gradually increased to 12-fold for 115 nA. Dose and DR depended on beam current, spot spacing, and arrangement and could be reproduced with 6.4% and 4.2% variations, respectively. Although FLASH was achieved and the largest field size that delivers FLASH DR was determined as 35 × 35 mm2, the current MIC has DR dependence, and users should measure dose and DR independently each time for their FLASH applications.
Author Notes
  • Serdar Charyyev, PhD, Department of Radiation Oncology, Stanford University School of Medicine, 875 Blake Wilbur Dr, Palo Alto, CA 94304, Tel: +1 (650)-724-9779, Email: charyyev@stanford.edu
Keywords
Research Categories
  • Health Sciences, Oncology

Tools

Relations

In Collection:

Items